Solta Medical Inc recalls VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the…
- Recall date
- June 1, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-3047-2017
- FDA classification
- Class II
- Brand / firm
- Solta Medical Inc
- Sold / distributed
- US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.
Why it was recalled
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Get recall alerts
Free email alert whenever Solta Medical Inc has a new recall — straight from official government data. Unsubscribe anytime.