Medical device recalls Moderate risk

Sorin Group USA, Inc. recalls AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8)…

Recall date
June 22, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2147-2016
FDA classification
Class II
Brand / firm
Sorin Group USA, Inc.
Sold / distributed
Worldwide Distribution: US (nationwide) to states of: LA, Ml, MO, OH, TN, and TX; and countries of: Italy and Canada.

Why it was recalled

Sorin Group USA is recalling Aortic Arch Cannulae (part numbers NA-55X7 and NA-55X8) because the distal end may be less than specified. The defective cannula are showing an angle that is about 15-20 degrees instead of approximately 45 degrees.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AORTIC ARCH CANNULA, Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass (part numbers NA-55X7 and NA-55X8) Aortic Arch Cannulae are designed to be used in the extracorporeal circuit during cardiopulmonary bypass surgery. The cannula consists of varying lengths of non-wire reinforced (flexible) polyvinyl chloride tubing that terminates in an angled tip. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.