Medical device recalls Moderate risk

Spacelabs Healthcare Inc recalls Spacelabs Healthcare CardioCall ECG Event Recorder, Models VS20, VS20x4, ST80, ST80x4, PCBAs PN # 670-1670-00 Product U…

Recall date
September 3, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0172-2016
FDA classification
Class II
Brand / firm
Spacelabs Healthcare Inc
Sold / distributed
Worldwide Distribution - US Distribution in the states of Nebraska, South Carolina, and Washington and in the following countries of AUSTRALIA, BELGIUM, BOLIVIA, CANADA, CHILE, DENMARK, GERMANY, GUATEMALA, HONG KONG, ITALY, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, QATAR, SINGAPORE, TURKEY, UNITE…

Why it was recalled

When the battery is inserted into the CardioCall ECG Event Recorder, the device beeps twice and is followed by a continuous ringing sound that only stops when battery is removed. The device has a self-test failure and will not operate normally.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Spacelabs Healthcare CardioCall ECG Event Recorder, Models VS20, VS20x4, ST80, ST80x4, PCBAs PN # 670-1670-00 Product Usage: The CardioCall event recorder is worn by the patient to record a short period of electrocardiogram which can be sent to his physician.