Medical device recalls Moderate risk

Spectranetics Corporation recalls Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in…

Recall date
June 6, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2863-2017
FDA classification
Class II
Brand / firm
Spectranetics Corporation
Sold / distributed
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, Hawaii and Puerto Rico Foreign distribution to Bahrain, Brazil, Canada, Italy, Japan, and Netherlands

Why it was recalled

Potential for flouroscopially visible marker band at the tip of the Rapid Exchange Turbo Elite Atherectomy Catheters to become loose or disengaged rom the catheter when the device is in use in an off-label manner.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Spectranetics Turbo-Elite Laser Atherectomy Catheter Peripheral Rapid Exchange Catheters Device is indicated for use in the treatment, including atherectomy, of infrainguinal stenosis and occlusions

Get recall alerts

Free email alert whenever Spectranetics Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Spectranetics Corporation