Medical device recalls Moderate risk

SPS Sterilization, Inc recalls Boston Scientific Wallstent Endoprosthesis Tracheobronchial Transhepatic Biliary Self-Expanding Stent, 12MM X 90MM | 75…

Recall date
March 7, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3035-2018
FDA classification
Class II
Brand / firm
SPS Sterilization, Inc
Sold / distributed
US distribution to AZ, CA, FL, OH, PA, TX, VA, WV

Why it was recalled

The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Boston Scientific Wallstent Endoprosthesis Tracheobronchial Transhepatic Biliary Self-Expanding Stent, 12MM X 90MM | 75CM L, resterilized.