Medical device recalls High risk

St Jude Medical Cardiac Rhythm Management Division recalls Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for lo…

Recall date
November 3, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0660-2016
FDA classification
Class I
Brand / firm
St Jude Medical Cardiac Rhythm Management Division
Sold / distributed
Worldwide Distribution US (nationwide), Belgium, Switzerland, Germany, France, United Kingdom, India, Italy, Japan, Luxembourg, The Netherlands, Saudi Arabia, and Sweden.

Why it was recalled

Leads may have been compromised during the manufacturing process, with cuts in the insulation layer of the SVC shock coil.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Cardioverter Defibrillator (ICD).