Sterilmed, Inc. recalls Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER AC…
- Recall date
- October 8, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0573-2026
- FDA classification
- Class II
- Brand / firm
- Sterilmed, Inc.
- Sold / distributed
- United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah
Why it was recalled
Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Component: N/A
Get recall alerts
Free email alert whenever Sterilmed, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Sterilmed, Inc.