Medical device recalls Moderate risk

Steris Corporation recalls ATLAS Loading Car

Recall date
December 26, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0824-2019
FDA classification
Class II
Brand / firm
Steris Corporation
Sold / distributed
Nationwide domestic distribution. Foreign distribution to Canada, Costa Rica, Indonesia, South Korea, and Taiwan.

Why it was recalled

The ATLAS loading car may not remain fully engaged with the latch on the transfer carriage when the loading car is empty, potentially resulting in the loading car unexpectedly sliding forward or falling to one side.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ATLAS Loading Car