Medical device recalls Moderate risk

Steris Corporation recalls Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry la…

Recall date
February 19, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1559-2015
FDA classification
Class II
Brand / firm
Steris Corporation
Sold / distributed
Worldwide Distribution - US Nationwide in the states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NH,NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI & WV and the countries of: Australia, Brazil, Canada, France…

Why it was recalled

Should the Caviwave Pro Ultrasonic Cleaning System experience a temporary interruption in electrical power during a processing cycle, the circulation pump may stop functioning. In addition, it is possible that the ultrasonic generator may stop working. The machine does not provide an alert notifying the user of this device malfunction.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Caviwave Pro Ultrasonic Cleaning System, Product Usage: To wash (remove tissue, blood and other contaminants) or dry laparoscopic and general surgical instruments prior to final disinfection and sterilization as required.

Get recall alerts

Free email alert whenever Steris Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Steris Corporation