Medical device recalls Moderate risk

Strata Skin Sciences, Inc. recalls XTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment of the skin…

Recall date
January 23, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1455-2017
FDA classification
Class II
Brand / firm
Strata Skin Sciences, Inc.
Sold / distributed
USA (nationwide) Distribution

Why it was recalled

When the Liquid Guide ( LLG) is not used according to the operator manual, the switch control wires may become loose and shorted. Shorted wires may lead to unintentional laser firing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

XTRAC Excimer Laser, Model No. AL8000, AL10000 The intended use is targeted UVB phototherapy for treatment of the skin conditions including psoriasis, vitiligo, atopic dermatitis, and leukoderma.