Stryker Communications recalls The Stryker Visum¿ LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures…
- Recall date
- February 28, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1699-2017
- FDA classification
- Class II
- Brand / firm
- Stryker Communications
- Sold / distributed
- Worldwide distribution - United States Nationwide - Australia, Canada, Great Britain, Japan, Netherlands and Puerto Rico
Why it was recalled
The potential hazard that can arise for the combination of the missing screws and the overloading by the end user is insufficient mount force that causes the equipment to fall.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The Stryker Visum¿ LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light. Affected part numbers are 0682001432 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/FP), 0682001286 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/FP), 0682001298 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P), 0682001299 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 5P/5P), 0682001300 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/5P/5P), 0682001433 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P), 0682001434 (EDS LIGHT SUSPENSION, CENTRAL AXIS, 9P/5P), 0682001435 (EDS LIGHT SUSPENSION, CENTRAL AXIS, FP/9P/5P) Product Usage: The Stryker Visum LED Surgical Lighting system is intended to illuminate the operative site during surgical procedures with high intensity light