Stryker Endoscopy recalls PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Mode…
- Recall date
- March 28, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1566-2016
- FDA classification
- Class II
- Brand / firm
- Stryker Endoscopy
- Sold / distributed
- Worldwide distribution. US nationwide including Puerto Rico, China, Hong Kong, India, Japan, Korea, Romania, Vietnam, Singapore, Taiwan, Netherlands, Peru, Australia, Canada, Mexico, Argentina, Chile, Poland, South Africa, and Guatemala.
Why it was recalled
When operating the device in high flow mode (40L/min, 15 mmHg) and connecting to a house gas inlet source, within seconds of insufflating the flow rate would drop to 0L/min and the red Check Gas Supply symbol would appear on the display, immediately suspending insufflation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PneumoSure XL High Flow Insufflator and PneumoSure High Flow Insufflator; High pressure units (HPU) (40L and 45L). Model Numbers: 620-040-610, 620-040-611, 620-040-613, 620-040-614, 620-040-600, 620-040-601, 620-040-602, 620-040-604, 105-210-684 The Pneumo Sure XL High Flow Insufflator is a device that serves to create a cavity by insufflating CO2 during diagnostics and/or therapeutic laparoscopy. The insufflator consists of a console, that is intended to be connected to a CO2 gas source, and has a tubeset inserted to transport CO2 gas from the insufflator to the connecting trocar.