Stryker Howmedica Osteonics Corp. recalls Kinemax Stem Extenders, 80 mm, Product Usage: Kinemax Cemented Stem Extenders are modular components designed to be mat…
- Recall date
- February 27, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1614-2015
- FDA classification
- Class II
- Brand / firm
- Stryker Howmedica Osteonics Corp.
- Sold / distributed
- US Nationwide distribution in the states of MA and RI.
Why it was recalled
Stryker received four customer complaints reporting that inside the box labeled part #6476-8-260 (Kinemax Stem Extenders, 80 mm) there was a different part #6476-8-250 (Kinemax Stem Extenders, 40 mm.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Kinemax Stem Extenders, 80 mm, Product Usage: Kinemax Cemented Stem Extenders are modular components designed to be mated with tibial baseplates or femoral components of the Modular Rotating Hinge (MRH), Scorpio Total Stabilizer (TS) or Duracon TS knee replacement systems. The Kinemax Cemented Stem Extenders are available in three different lengths, 40 mm (Part #6476-8-250), 80mm (Part #6476-8-260) and 155mm (Part #6476-8-270). The 40 mm cemented stem extender is not indicated for use in the MRH system; however, it is indicated for use with the Scorpio TS and Duracon TS systems (LRKS/G Rev. 2). Both the 80mm and 155mm stem extenders are indicated for use in all three systems.
Get recall alerts
Free email alert whenever Stryker Howmedica Osteonics Corp. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Stryker Howmedica Osteonics Corp.