Medical device recalls Moderate risk

Stryker Orthopaedics Patella Assembly Instrument recalled over sterility concerns

Recall date
August 1, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Stryker Howmedica Osteonics Corp. recalls Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio…
Recall number
Z-2789-2016
FDA classification
Class II
Brand / firm
Stryker Howmedica Osteonics Corp.
Sold / distributed
US Nationwide and Internationally

Why it was recalled

Upon failure of the press-fit, the clamping sub components and pins could disassociate from the Scorpio Patella Assembly Instrument.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Stryker Orthopaedics Patella Assembly Instrument, REF 3182-1000, Non-Sterile, NDC (01) 07613327215854 (10) The Scorpio Patella Assembly Instrument is used during total knee replacement surgery for the implantation of Scorpio Patella implant components. The device functions to fully seat the implant onto the bone and secure it in position while the surgeon removes any excess cement and when cement polymerization occurs.