Stryker Howmedica Osteonics product recalled over sterility concerns
- Recall date
- June 23, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Stryker Howmedica Osteonics Corp. recalls Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement o…
- Recall number
- Z-2390-2016
- FDA classification
- Class II
- Brand / firm
- Stryker Howmedica Osteonics Corp.
- Sold / distributed
- Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT
Why it was recalled
It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants
Get recall alerts
Free email alert whenever Stryker Howmedica Osteonics Corp. has a new recall — straight from official government data. Unsubscribe anytime.