Medical device recalls Moderate risk

Stryker Howmedica Osteonics product recalled over sterility concerns

Recall date
June 23, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Stryker Howmedica Osteonics Corp. recalls Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement o…
Recall number
Z-2390-2016
FDA classification
Class II
Brand / firm
Stryker Howmedica Osteonics Corp.
Sold / distributed
Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT

Why it was recalled

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants

Get recall alerts

Free email alert whenever Stryker Howmedica Osteonics Corp. has a new recall — straight from official government data. Unsubscribe anytime.