Stryker Instruments Div. of Stryker Corporation recalls Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long co…
- Recall date
- January 13, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1098-2020
- FDA classification
- Class II
- Brand / firm
- Stryker Instruments Div. of Stryker Corporation
- Sold / distributed
- Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.
Why it was recalled
Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332