Medical device recalls Moderate risk

Stryker Instruments Div. of Stryker Corporation recalls Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long co…

Recall date
January 13, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1098-2020
FDA classification
Class II
Brand / firm
Stryker Instruments Div. of Stryker Corporation
Sold / distributed
Domestic distribution nationwide. Foreign distribution to Netherlands and Australia.

Why it was recalled

Hairline fractures in a component could allow electrical current to arc out of the device, resulting in a burn to the patient and/or user.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Telescopic Smoke Evac Pencil, PB, Coated, Product No. SHKTSPCL. The product is a Bovie pencil that is 75 inches long compressed, with a telescoping shaft. UDI GTIN 37613327466332

Get recall alerts

Free email alert whenever Stryker Instruments Div. of Stryker Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Stryker Instruments Div. of Stryker Corporation