Stryker Instruments Div. of Stryker Corporation recalls The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software an…
- Recall date
- October 16, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1216-2018
- FDA classification
- Class II
- Brand / firm
- Stryker Instruments Div. of Stryker Corporation
- Sold / distributed
- International , US and US Govt.
Why it was recalled
Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are also helpful in determining the appropriate pedicle screw length.