Stryker Leibinger GmbH & Co. KG recalls 8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solutio…
- Recall date
- April 4, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1885-2024
- FDA classification
- Class II
- Brand / firm
- Stryker Leibinger GmbH & Co. KG
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
8000-021-002, Scopis¿ ENT Software with TGS¿ with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution for navigating functional endoscopic sinus surgery (FESS) offering surgeons highly advanced image guidance and visualization capabilities in a single system. The Stryker Electromagnetic Navigation Unit (8000-010-003) powered by Scopis ENT Software with TGS (8000-021-002), allows the analysis and identification of sinus cells in the complex patient anatomy and planning of the natural drainage pathways through the sinus cavity. During surgery, the planned pathways are overlaid in real-time onto the endoscopic image, providing a unique Scopis augmented reality technology. Guidance of endoscopic instruments may help perform a minimally invasive, accurate and selective surgery.
Get recall alerts
Free email alert whenever Stryker Leibinger GmbH & Co. KG has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Stryker Leibinger GmbH & Co. KG