Stryker Neurovascular recalls Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils ar…

Recall date
June 6, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2183-2016
FDA classification
Class II
Brand / firm
Stryker Neurovascular
Sold / distributed
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamah…

Why it was recalled

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.