Medical device recalls Moderate risk

Stryker Sustainability Solutions recalls Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134302, Steri…

Recall date
January 2, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0917-2018
FDA classification
Class II
Brand / firm
Stryker Sustainability Solutions
Sold / distributed
USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.

Why it was recalled

An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134302, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.