Medical device recalls Moderate risk

Synthes (USA) Products LLC recalls Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set

Recall date
March 5, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2245-2018
FDA classification
Class II
Brand / firm
Synthes (USA) Products LLC
Sold / distributed
Distributed to accounts in CA, IL, IN, KS, LA, MD, NC, NE, NJ, OR, TX, and VA. Foreign distribution to Canada.

Why it was recalled

Affected screw gauges incorrectly measure screws 5.41mm shorter than the actual overall screw length.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Part no. 690.375, Graphic Case F/6.5mm & 7.3mm Combined Cannulated Screw Set

Get recall alerts

Free email alert whenever Synthes (USA) Products LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Synthes (USA) Products LLC