Medical device recalls Moderate risk

Synthes (USA) Products LLC recalls Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Ne…

Recall date
February 23, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1327-2016
FDA classification
Class II
Brand / firm
Synthes (USA) Products LLC
Sold / distributed
Worldwide Distribution - US Nationwide in the states of MA, VA, TN, GA, WA, IA, NJ, TN, PA, NY, CO, CT, VT, and the countries of Canada and Switzerland.

Why it was recalled

Out of specification at the thread. The thread height of these screws is under-sized. The cross section of the affected area is less than that of conforming screws such that the thread height of the screws is undersized.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Titanium Low Profile Neuro Screw, Self Drilling, 3mm, part numbers: 400.833, 400.833E Product Usage: The Low Profile Neuro Plating System is intended for use in selective trauma of the midface and craniofacial skeleton; craniofacial surgery; reconstructive procedures; and selective orthognathic surgery of the maxilla and chin.

Get recall alerts

Free email alert whenever Synthes (USA) Products LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Synthes (USA) Products LLC