TAS Medical Inc recalls TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
- Recall date
- January 27, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1294-2025
- FDA classification
- Class II
- Brand / firm
- TAS Medical Inc
- Sold / distributed
- US Nationwide distribution in the states of LA, OK, TN.
Why it was recalled
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
TISSUE APPROXIMATION SYSTEM (TAS), Models T-4000, T-5000, T-LAP
Get recall alerts
Free email alert whenever TAS Medical Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: TAS Medical Inc