Medical device recalls Moderate risk

Tec Com Gmbh recalls TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product…

Recall date
August 14, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0362-2020
FDA classification
Class II
Brand / firm
Tec Com Gmbh
Sold / distributed
US Nationwide distribution in the state of UT.

Why it was recalled

The action is being initiated due to potential installation of faulty peltier-elements within the thermo-regulation system. Failure of the peltier-element would report a system failure and no longer be able to guarantee temperature control at full power.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

TECOTHERM NEO, 100-130/200-240 VAC, 50-60 Hz/ max. 350VA, Class I, Schutzklasse I IP 20, UDI: 04260498580002 - Product Usage: The Thermo-Regulation System TECOTHERM NEO is designed for controlled cold & heat treatment procedures. By means of a mattress, cold and heat is provided to the patient, depending on the therapy objective.

Get recall alerts

Free email alert whenever Tec Com Gmbh has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Tec Com Gmbh