Medical device recalls High risk

Teleflex Medical recalls Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07K…

Recall date
May 10, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1912-2019
FDA classification
Class I
Brand / firm
Teleflex Medical
Sold / distributed
US Nationwide Distribution in the states of AL, TX

Why it was recalled

Reported complaints of cracks being observed, prior to use, on swivel wye adaptors of the Neonatal ConchaSmart Breathing Circuit.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Teleflex HUDSON RDI NEONATAL ConchaSmart Breathing Circuit with Dual Heated Limb and ConchaSmart Column: a) REF 870-07KIT b) REF 870-09KIT Product Usage: The Neonatal ConchaSmart breathing circuits are intended to deliver humidified breathing gases for administration to an infant/neonate patient. These gases may be delivered by nasal prongs using the CareFusion Infant Flow system and the single limb circuit, or through bypassing upper airways, for example through the use of an endotracheal tube with the dual limb configuration.