Medical device recalls Moderate risk

Terumo Cardiovascular Systems Corporation recalls CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial…

Recall date
May 28, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2018-2025
FDA classification
Class II
Brand / firm
Terumo Cardiovascular Systems Corporation
Sold / distributed
US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.

Why it was recalled

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Get recall alerts

Free email alert whenever Terumo Cardiovascular Systems Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Terumo Cardiovascular Systems Corporation