Terumo Cardiovascular Systems Corporation recalls HX2" Temperature Management Systems; The Terumo¿ HX2" Temperature Management System provides temperature control of two…
- Recall date
- June 8, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2207-2016
- FDA classification
- Class II
- Brand / firm
- Terumo Cardiovascular Systems Corporation
- Sold / distributed
- Domestic: AK, AL, AR, AZ, CA , CO, CT, DC , DE , FL, GA, HI, IA, ID, IL, IN, KS, KY,LA , MA, MD, ME, MI , MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV , NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY Foreign: Argentina, Australia, Belgium,Brazil, Canada, Chile, China, Colombia, Co…
Why it was recalled
Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
HX2" Temperature Management Systems; The Terumo¿ HX2" Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery. The system consists of a water tank, circulating pumps, heater manifolds, mercury free temperature sensors, water detectors, mixing valves and a tank divider which is provided to partition the tank into two separate channels (Left and Right).The system has the capacity to circulate water at a rate of up to 6.5 gal.min (25 L/min) with no load connected. The system is capable of heating and cooling for a single channel or for both channels. Product Usage: HX2 Temperature Management Systems is indicated for use to supply temperature regulated water to heat exchangers during open heart surgery.
Get recall alerts
Free email alert whenever Terumo Cardiovascular Systems Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Terumo Cardiovascular Systems Corporation