Medical device recalls Moderate risk

Terumo Cardiovascular Systems Corporation recalls Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly c…

Recall date
February 6, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1116-2015
FDA classification
Class II
Brand / firm
Terumo Cardiovascular Systems Corporation
Sold / distributed
US (nationwide) including the states of AR, CA, CO, CT, FL, IA, IL, KS, KY, MD, MI, MO, MS, NY, OH, PA, TX, UT, WA, and WI and the country of Japan.

Why it was recalled

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Terumo HX2 Temperature Management System provides temperature control of two independent water circuits that directly controls the temperature of patient blood and cardioplegia solution during cardiovascular surgery.