Medical device recalls Moderate risk

The Anspach Effort, Inc. recalls ANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.

Recall date
April 13, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2282-2015
FDA classification
Class II
Brand / firm
The Anspach Effort, Inc.
Sold / distributed
Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.

Why it was recalled

G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ANSPACH - Extra Coarse Diamond Drum, 6.8MM x 10.6MM REF L-68DRMDX-G1 Cutting shaping bone including spine and cranium.