Medical device recalls Moderate risk

The Anspach Effort, Inc. recalls Pediatric Craniotome Ref:CRANI-P-G1

Recall date
December 7, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0924-2023
FDA classification
Class II
Brand / firm
The Anspach Effort, Inc.
Sold / distributed
US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of China Thailand Angola Austria Belgium Croatia Cyprus Czech Republic Denmark Equatorial Guinea Finlan…

Why it was recalled

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Pediatric Craniotome Ref:CRANI-P-G1

Get recall alerts

Free email alert whenever The Anspach Effort, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: The Anspach Effort, Inc.