Medical device recalls Moderate risk

The Binding Site Group, Ltd. recalls Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum…

Recall date
July 25, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0477-2019
FDA classification
Class II
Brand / firm
The Binding Site Group, Ltd.
Sold / distributed
U.S.: IL, WV, NE, SD, NY, IN, WA, MD, NJ, PA, NC, AR, FL, VA

Why it was recalled

After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Binding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgM antiserum, six calibrators and two controls. The calibrators and controls are based on stabilized human serum. All components contain sodium azide as a preservative and are provided in ready-to-use liquid format.