Medical device recalls Moderate risk

The Binding Site Group, Ltd. recalls Freelite Human Lambda Free Kit for use on SPAPLUS Product Code: LK018.S, LK018.10S Freelite aids the detection and moni…

Recall date
October 28, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1053-2017
FDA classification
Class II
Brand / firm
The Binding Site Group, Ltd.
Sold / distributed
Worldwide Distribution - U.S. (nationwide) and Internationally to the following; Canada, Australia.

Why it was recalled

A change in the calibration curve causing an increase in false prozone flags.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Freelite Human Lambda Free Kit for use on SPAPLUS Product Code: LK018.S, LK018.10S Freelite aids the detection and monitoring of Multiple Myeloma and related diseases

Get recall alerts

Free email alert whenever The Binding Site Group, Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: The Binding Site Group, Ltd.