Medical device recalls Moderate risk

The Binding Site Group, Ltd. recalls OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565

Recall date
July 29, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1327-2020
FDA classification
Class II
Brand / firm
The Binding Site Group, Ltd.
Sold / distributed
CA, PA, FL, AR, SD, MA, WV, AL, GA, IL, MN, MD, AZ, NC, TN, WA, MI, NY, KY, SC, IN, OR, NE, OH, TX, VT, WA, UT, IA,

Why it was recalled

There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting outside the specified ranges. This could result in potential false patient results which could cause a delay until repeat testing is performed.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565

Get recall alerts

Free email alert whenever The Binding Site Group, Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: The Binding Site Group, Ltd.