The Binding Site Group, Ltd. recalls OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565
- Recall date
- July 29, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1327-2020
- FDA classification
- Class II
- Brand / firm
- The Binding Site Group, Ltd.
- Sold / distributed
- CA, PA, FL, AR, SD, MA, WV, AL, GA, IL, MN, MD, AZ, NC, TN, WA, MI, NY, KY, SC, IN, OR, NE, OH, TX, VT, WA, UT, IA,
Why it was recalled
There is a potential that users may experience difficulty validating the calibration of the kit. This is indicated by kit control values reporting outside the specified ranges. This could result in potential false patient results which could cause a delay until repeat testing is performed.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
OPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565