Thermo Fisher Scientific (Asheville) LLC recalls Thermo Scientific General Purpose Dynamic Bath Precision 2L-Shallow GP Bath Model Number: TSGP2S
- Recall date
- April 2, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1870-2019
- FDA classification
- Class II
- Brand / firm
- Thermo Fisher Scientific (Asheville) LLC
- Sold / distributed
- CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Why it was recalled
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Thermo Scientific General Purpose Dynamic Bath Precision 2L-Shallow GP Bath Model Number: TSGP2S