- Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAY…
Why it was recalled
Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System