Medical device recalls High risk

Thornhill Research Inc recalls MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transpor…

Recall date
November 19, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0587-2016
FDA classification
Class I
Brand / firm
Thornhill Research Inc
Sold / distributed
Worldwide Distribution: US distribution in state of: Virginia and to the country of: Singapore.

Why it was recalled

Batteries found unresponsive after storage, not recognized by main device, and are perpetually enabled.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

MOVES System Lithium Polymer Batteries The MOVES System is a portable computer controlled electrically powered transport emergency ventilator intended to provide continuous or intermittent ventilatory support for the care of adults who require ventilation.