Medical device recalls Moderate risk

Tollos recalls Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600,…

Recall date
October 24, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0245-2023
FDA classification
Class II
Brand / firm
Tollos
Sold / distributed
Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.

Why it was recalled

Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000