Medical device recalls Moderate risk

Toshiba American Medical Systems Inc recalls Toshiba America Medical Systems, Inc . Aquilion LB TSX-201A/2, 3 CT Scanner

Recall date
April 25, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0048-2017
FDA classification
Class II
Brand / firm
Toshiba American Medical Systems Inc
Sold / distributed
US Distribution to the states of: PA, IA, MA, SD, MI, LA, KY, NC, NY, AR, IN, WI, OH, TY, FL, WV, WA, CT, MS, AL, ND, CO, HI, NO, MN, ID, OH, GA, MT, LA, NM, VT, TN, CA, NH, NJ, NE, SC and KS.

Why it was recalled

It has been found that if the scan range extends beyond the maximum Field of View (FOV), a reconstruction operation error may occur. As a result, scanning may be interrupted and rebooting of the system may be necessary. Please note that this issue has not occurred in the U.S.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Toshiba America Medical Systems, Inc . Aquilion LB TSX-201A/2, 3 CT Scanner