Tosoh Bioscience Inc recalls AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable…
- Recall date
- September 28, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0599-2019
- FDA classification
- Class II
- Brand / firm
- Tosoh Bioscience Inc
- Sold / distributed
- US Nationwide distribution and International distribution to Colombia, Chile, Honduras, Cayman Islands, Panama, Costa Rica, Guatemala, Venezuela, Uruguay, Peru, and Ecuador.
Why it was recalled
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
AIA-2000 Analyzer, Product code 022100, 022100R, 022101, 022101R The AIA-2000 analyzer and AIA-900 analyzer are capable of performing three methods of immunoassay: an immunoenzymetric (IEMA) or sandwich immunoassay, a competitive binding (EIA) immunoassay, and a two-step immunoenzymetric immunoassay.
Get recall alerts
Free email alert whenever Tosoh Bioscience Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Tosoh Bioscience Inc