Medical device recalls Moderate risk

Tosoh Bioscience Inc recalls Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitativ…

Recall date
February 23, 2012
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1187-2021
FDA classification
Class II
Brand / firm
Tosoh Bioscience Inc
Sold / distributed
US Nationwide distribution in the states of CA, FL, GA, NY, OH & OR.

Why it was recalled

Some combinations of Calibrator lots and AlA PACKS resulted in Calibrator #6 not giving a value, making the calibration unacceptable.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Thyroglobulin Antibody (TgAB) Calibrator - Product Usage: intended for IN VITRO DIAGNOSTIC USE ONLY for the quantitative measurement of anti-thyroglobulin antibodies (TgAb) in human serum on specific TOSOH AIA System analyzers. Measurements obtained by this device aid in the diagnosis of autoimmune thyroid disorders, such as Graves disease and Hashimotos disease. Comment to IU: Please note that the Calibrator set is not part of the IU but required for the assay.

Get recall alerts

Free email alert whenever Tosoh Bioscience Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Tosoh Bioscience Inc