Medical device recalls Moderate risk

Tosoh Bioscience Inc recalls Tosoh AIA-900 immunoassay Analyzer

Recall date
September 28, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0485-2019
FDA classification
Class II
Brand / firm
Tosoh Bioscience Inc
Sold / distributed
Distributed to accounts in GA and NY.

Why it was recalled

A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released software version 2.51 (or later) to fix the problem. This failure is mitigated by the hardware design. The consequence would be a slight delay in testing patient sample or reporting patient result.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Tosoh AIA-900 immunoassay Analyzer

Get recall alerts

Free email alert whenever Tosoh Bioscience Inc has a new recall — straight from official government data. Unsubscribe anytime.