Turncare, Inc recalls Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Versio…
- Recall date
- August 10, 2023
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1968-2025
- FDA classification
- Class II
- Brand / firm
- Turncare, Inc
- Sold / distributed
- US Nationwide distribution in the states of Texas, Illinois, Pennsylvania, Florida, Connecticut.
Why it was recalled
A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brand Name: Guardian 2 System Product Name: Guardian System 2 Controller Model/Catalog Number: GS-2.5-C Software Version: 2.5.4-69 (Fix)