Varian Medical Systems Inc recalls The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical…
- Recall date
- May 18, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1849-2015
- FDA classification
- Class II
- Brand / firm
- Varian Medical Systems Inc
- Sold / distributed
- Worldwide Distribution.
Why it was recalled
After an application freeze and restart of the VariSource iX series by power cycling, the Partial Fraction generated by the system will not be correct. The application freeze issue affects only the systems equipped with magnetic Hard Disk Drives (HDD).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The VariSource iX series afterloader systems are computer controlled remote electro/mechanical systems used for medical purposes, for placing a cable incorporating an irradiated iridium seed internally or close by a malignant tumor or tumor bed in a practice known as brachytherapy.