Ventana Medical Systems, Inc. recalls VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD P…
- Recall date
- November 18, 2022
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0907-2023
- FDA classification
- Class II
- Brand / firm
- Ventana Medical Systems, Inc.
- Sold / distributed
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Cz…
Why it was recalled
Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
Get recall alerts
Free email alert whenever Ventana Medical Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ventana Medical Systems, Inc.