Villa Sistemi Medicali S.P.A. recalls Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray sys…
- Recall date
- November 4, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0764-2020
- FDA classification
- Class II
- Brand / firm
- Villa Sistemi Medicali S.P.A.
- Sold / distributed
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
Why it was recalled
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The movements of the table and the tube stand are motorized and remotely controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.