Medical device recalls Moderate risk

Vision RT Ltd recalls AlignRT¿ Plus [private label-OSMS] K123371: The AlignRT Plus system is indicated for use to position and monitor patien…

Recall date
June 30, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2667-2016
FDA classification
Class II
Brand / firm
Vision RT Ltd
Sold / distributed
Product distributed in US. The foreign countries where the product was distributed are as follows: Canada, Germany, Japan, Sweden, Switzerland, Taiwan, and United Kingdom. No product was distributed to US government accounts.

Why it was recalled

Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up for the Optical Surface Monitoring System [OSMS]. 1. CRM 3194, No Serious Injury, No MDR, Aware date: Dec 11, 2015 2. CRM 3321, No Patient Involved, No MDR, Aware date: Jan 19, 2016 3. CRM 4570, No Serious Injury, No MDR, Aware date: March 15, 2016 No patient harm was reported in any of these cases.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

AlignRT¿ Plus [private label-OSMS] K123371: The AlignRT Plus system is indicated for use to position and monitor patients relative to the prescribed treatment isocentre, and to withhold the beam automatically during radiation delivery.