Vision RT Ltd recalls TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The TrueBeam and Edge Systems a…
- Recall date
- June 30, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2668-2016
- FDA classification
- Class II
- Brand / firm
- Vision RT Ltd
- Sold / distributed
- Product distributed in US. The foreign countries where the product was distributed are as follows: Canada, Germany, Japan, Sweden, Switzerland, Taiwan, and United Kingdom. No product was distributed to US government accounts.
Why it was recalled
Vision RT Ltd received three (3) reports from different sites that discovered the following issue, unintended changes were made to planned couch parameters, specifically the couch rotation parameter, during patient set-up for the Optical Surface Monitoring System [OSMS]. 1. CRM 3194, No Serious Injury, No MDR, Aware date: Dec 11, 2015 2. CRM 3321, No Patient Involved, No MDR, Aware date: Jan 19, 2016 3. CRM 4570, No Serious Injury, No MDR, Aware date: March 15, 2016 No patient harm was reported in any of these cases.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
TrueBeam¿ Radiotherapy Delivery System and EDGE" Radiotherapy Delivery System, K140528. The TrueBeam and Edge Systems are intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation therapy is indicated for adults and pediatric patients.
Get recall alerts
Free email alert whenever Vision RT Ltd has a new recall — straight from official government data. Unsubscribe anytime.