Medical device recalls Moderate risk

Volcano Corp recalls Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 8…

Recall date
June 13, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2221-2025
FDA classification
Class II
Brand / firm
Volcano Corp
Sold / distributed
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries…

Why it was recalled

Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Volcano Visions Digital IVUS Catheter: PV.014P (Platinum), REF:85910P; PV.014P RX, REF: 014R; and PV.018, REF: 86700, 86700J used with Instructions for use.

Get recall alerts

Free email alert whenever Volcano Corp has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Volcano Corp