W L Gore & Associates, Inc. recalls GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, E…
- Recall date
- August 26, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0036-2026
- FDA classification
- Class II
- Brand / firm
- W L Gore & Associates, Inc.
- Sold / distributed
- US: OR, AL, OH, MD, VA, CA, FL, NC, OK, MA, MO, AZ, TX, IL, GA, PA, NH, HI, MN, MS, KY, ID, LA, NJ, TN, DE, IN, MI, WI, NY, DC, SC, CT, NE, IA, WA, AR, NV, WV, AK, UT, ND, KS, CO, ME, MT, NM, SD, PR, WY, RI. OUS: PANAMA, CANADA, CHINA, TAIWAN, GERMANY, SLOVENIA, NORWAY, UNITED KINGDOM, ITALY, SPAIN…
Why it was recalled
Reports of vascular graft delamination, which may lead to reduced access flow; cannulation difficulties; bleeding or bruising; and harms related to reintervention procedures, so Instructions for Use will be updated to modify existing warning to 1) further clarify techniques that may contribute to the risk of delamination, 2) Addition of delamination to device-related adverse event section.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
GORE ACUSEAL Vascular Graft, REF: ECH050020J, ECH050020W, ECH050050J, ECH050050W, ECH060010A, ECH060020A, ECH060020J, ECH060020W, ECH060040, ECH060040A, ECH060040W, ECH060050A, ECH060050J, ECH460045A, ECH460045J, ECH470045, ECH470045A; used with Instructions for Use.