Waismed Ltd recalls NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle…

Recall date
August 18, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2680-2016
FDA classification
Class II
Brand / firm
Waismed Ltd
Sold / distributed
WI, TX, CA, LA and the following countries: UK, Spain, Poland, Israel, Czech Republic, France, Belgium, Denmark, Italy, Cyprus, Austria, Greece, Slovenia, Australia, Germany, Estonia, New Zealand, Switzerland, Costa Rica, Guatemala, Canada, Malaysia, Poland, Singapore, Lithuania, Romania

Why it was recalled

WaisMed received a report in which during the operation of the NIO-A, the needle was not released from the device as expected. The device was placed on the floor and after several minutes the needle was released spontaneously. NIO-A and the NIO-P models have the same activation mechanism. However, none of the affected lots of the NIO-A are distributed in the US, and thus the company is only recalling a single lot of NIO-P in the US.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

NIO-A / NIO-P is an instant IO access device comprises housing with spring load mechanism that injects a trocar needle into the bone marrow of a patient and allows connection of syringe or IV administration set. The NIO-A is intended for adults above 12 years old and the NIO-P is intended for pediatric patients 3 - 12 years of age.